To check the third pressure levels. “RUN” will appear on LCD and
pressures is checked at 0/120/90/60 mmHg points. If the pressure
difference at 0/120/90/60 mmHg is within ± 3mmHg, the calibrator will
rapid exhaust to 0 mmHg, then go on checking 3rd points. If "FAIL"
appears, the faulty pressure points will be shown (both the default
pressure check point and the actual pressure measured will appear
alternatively.).
If "FAIL" happens, press the "Auto calibration" key for re-calibration.
After this process is complete, press the "Readout" key for double
check.
check point on
Calibrator
check point from Blood
Pressure Monitor
If deviation under ±3mmHg
If deviation over ±3mmHg
Automatic Check 4th
Pressure Points
Calibrator Screen
Blood Pressure
Monitor Screen
Calibrator Screen
Blood Pressure
Monitor Screen
Press Auto Calibration Mode
(follow instructions on Page 18)
Calibrator Screen
Blood Pressure
Monitor Screen
Checking the pressures between calibrator and
monitor. If the deviation of Second Pressure Points
is under ±3mmHg, the calibrator would
automatic check Third Pressure Points
(120/90/60 mmHg)
Second
Pressure
Points
If deviation between calibrator and
monitor is under ±3mmHg
Third
Pressure
Points