To check the third pressure levels. “RUN” will appear on LCD and 
pressures is checked at 0/120/90/60 mmHg points. If the pressure 
difference at 0/120/90/60 mmHg is within ± 3mmHg, the calibrator will 
rapid exhaust to 0 mmHg, then go on checking 3rd points. If "FAIL" 
appears, the faulty pressure points will be shown (both the default 
pressure check point and the actual pressure measured will appear 
alternatively.).

If "FAIL" happens, press the "Auto calibration" key for re-calibration. 
After this process is complete, press the "Readout" key for double 
check.

check point on 

Calibrator

check point from Blood 

Pressure Monitor

If deviation under ±3mmHg

If deviation over ±3mmHg

Automatic Check 4th 

Pressure Points

Calibrator Screen

Blood Pressure 

Monitor Screen

Calibrator Screen

Blood Pressure 

Monitor Screen

Press Auto Calibration Mode 

(follow instructions on Page 18)

Calibrator Screen

Blood Pressure 

Monitor Screen

Checking the pressures between calibrator and 

monitor. If the deviation of Second Pressure Points  

is under ±3mmHg, the calibrator would  

automatic check Third Pressure Points  

(120/90/60 mmHg)

Second 

Pressure 

Points

If deviation between calibrator and 

monitor is under ±3mmHg

Third 

Pressure 

Points

Содержание CalPro

Страница 1: ...Instruction Manual Model CalPro Mobile Calibration System for Blood Pressure Monitor...

Страница 2: ...and conforms to the provi sions of the Electromagnetic Compatibility EMC Directive 2014 30 EU Low Voltage Directive LVD 2014 35 EU and EN IEC 61010 1 2010 Safe ty requirements for electrical equipment...

Страница 3: ...key ON OFF key Readout key Auto Calibration key Zero key Deflate key Inflate key Check Count key Clear Count key Set Count key Air tube socket Data Link Port AC Adapter Jack Data Link Cable Air Tube a...

Страница 4: ...sing the device Connection Methods Upper Arm Blood Pressure Monitor 1 Connect the tube between the calibrator and the monitor to the air tube socket 2 Then connect the data link cable between the cali...

Страница 5: ...ck count Clear count Set count Operation Procedures 1 Data link status operation Before operation make sure the air tube and the data link cable are securely connected to the calibrator and the blood...

Страница 6: ...ssure points will be shown both the default pressure check point and the actual pressure measured will appear alternatively If FAIL appears press the Auto calibration key for re calibration After this...

Страница 7: ...s complete press the Readout key for double check Checking the pressures between calibrator and monitor If the deviation of First Pressure Points is under 3mmHg the calibrator will automatically check...

Страница 8: ...s complete press the Readout key for double check check point on Calibrator check point from Blood Pressure Monitor If deviation under 3mmHg If deviation over 3mmHg Automatic Check 4th Pressure Points...

Страница 9: ...dout key for double check check point on Calibrator check point from Blood Pressure Monitor If deviation under 3mmHg If deviation over 3mmHg Readout Check is OK Calibrator Screen Blood Pressure Monito...

Страница 10: ...successful PASS will appear Otherwise FAIL will appear If FAIL continues to appear return the monitor to the manufacturer for service mmHg Auto calibration mmHg Auto calibration mmHg Auto calibration...

Страница 11: ...cessful PASS will appear and the number of uses appears Otherwise FAIL will appear 5 Measurement Count Clearance Mode Press Clear count to start RUN appears on LCD and the measurement cycles count num...

Страница 12: ...essure checking of Rossmax digital manometer Before entering this mode set the monitor to Test Mode To set your monitor to Test Mode 7 1 Remove 1 battery If using adapter as a power source disconnect...

Страница 13: ...key the pressure will decrease 1 mmHg accordingly i n fl a t e i n fl a t e d e fl a t e d e fl a t e i n fl a t e i n fl a t e d e fl a t e d e fl a t e i n fl a t e i n fl a t e d e fl a t e d e fl...

Страница 14: ...re Monitor and perform the auto calibration mode again ERR Connection Error Turn off the Blood Pressure Monitor disconnect and reconnect the data link cable turn on the Blood Pressure Monitor and oper...

Страница 15: ...14 F 140 F 10 90 RH max Points of Pressure Programming towards Blood Pressure Monitor 0 300 200 100 mmHg Points of Pressure Inspection towards Blood Pressure Monitor 0 30 60 90 120 150 180 210 240 27...

Страница 16: ...oes not cover damage accidents or non compliance with the instruction manual Please contact your local seller dealer or www rossmax com Customer Name _____________________________________________ Addr...

Страница 17: ...OBM_IB_CalPro_ EN_SW_ver1830 IN0CALPRO0000000XX Manufacturer Rossmax Swiss GmbH Tramstrasse 16 CH 9442 Berneck Switzerland www rossmax com...

Отзывы: