26
27
Error
Messages
Root Cause
Correction
FAIL
Connection Error
Turn off the Blood Pressure
Monitor, disconnect and reconnect
the data link cable, turn on the
Blood Pressure Monitor and
operate again. If "FAIL" continues
to appear, return the monitor to
the manufacturer.
Data Error (the pressure
difference between the
Blood Pressure Monitor
and the calibrator is over
3mmHg)
Turn off the Blood Pressure
Monitor, disconnect and reconnect
the data link cable, turn on the
Blood Pressure Monitor and
perform the auto calibration mode
again.
ERR
Connection Error
Turn off the Blood Pressure
Monitor, disconnect and reconnect
the data link cable, turn on the
Blood Pressure Monitor and
operate again. If "ERR" continues
to appear, return the calibrator to
the manufacturer.
Change "mmHg" to "kPa"
When the calibrator is off, press and hold "Auto calibration" key,
then press the "ON/OFF" key to turn on the calibrator, "mmHg" will
changed to "kPa" (release "Auto calibration" key), the calibrator will
show "kPa" on the LCD. Repeat each time, after powering off as the
calibrator default in "mmHg".
Error Messages
kPa mmHg
Readout
Auto calibration
Pressure gauge
Check
count
Clear
count
Set
count
+