TRM

 (7-term) 

 

Thru-Reflect-Match 

 
The male and female Short standards of this kit are designed with the same offset length. So 
the  TRM  calibration  technique  can  be  used.  It  requires  high  reflects  with  a  symmetrical 
reflection  during  VNA  calibration  on  both  test  ports.  TRM  needs  less  connections  during 
calibration than OSLT and UOSM. 
 

UOSL (SOLR)

 (7-term) 

Unknown-Open-Short-Load (Short-Open-Load- Reciprocal) 

 
The  UOSM  technique  is  similar  to  the  OSLT  calibration  but  doesn’t  need  well-known 
adapters. Even a low loss Device under Test (DUT) can be sufficient as Thru standard. The 
UOSM technique is particularly well suited for calibration involving different connector types 
at the test ports and is most effective when data based standards are available. 
 
In cases when the two VNA test ports have different connector systems installed, e.g. RPC-N 
50 

 and RPC-3.50 UOSM can handle this situation. A second calibration kit is needed. 

 

Calibration Procedure 

 
The procedures to perform a VNA calibration are very different between the VNA models. Modern 
VNA often guide you thru the settings and actions. For details please consult the manual of your 
specific VNA. 
 

4.2 Calibration Check 

 

It is advisable to check the calibration after error correction has been performed, particularly when 
test port cables and adapters have been used. 
 
During the calibration 

  Calibration standards can easily be mixed up 

  Connections may not be tightened correctly 

 
There might be defects 

  VNA, e.g. mechanically overloaded test port 

  Instable test port cables 

  Mechanically overloaded load 

  Bent or broken contact fingers 

 
With  a  Verification  Kit,  reflection  and  transmission  measurements  can  be  checked  in  a  wide 
frequency range for DUTs with low and high reflections and low and medium insertion loss. Actual 
measurement  values  can  be  compared  to  reference  values  delivered  with  the  Verification  Kit 
standards. This should be done from time to time to identify defects.  
 
Air  Lines,  Precision  Adapters  and  Mismatch  Standards  can  also  be  used  as  a  faster  method  for 
more frequent checks. 

 

 

Содержание 05CK001-150

Страница 1: ...uenztechnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 17 Dieses Dokument ist urheberrechtlich gesch tzt This document is...

Страница 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installa...

Страница 3: ...ibration kit used Contributing factors are noise linearity and temperature stability of the VNA transmission and reflection stability of a test port cable as well as operator experience and care VNAs...

Страница 4: ...C10S3 1 Calibration load jack 05K150 C10S3 1 Calibration adaptor plug plug 05S121 S20S3 1 Calibration adaptor jack jack 05K121 K20S3 1 Combi wrench 53W011 000 13 mm 14 mm 15 mm 16 mm 1 Torque wrench...

Страница 5: ...n for installing Standard Definitions on a Vector Network Analyzer Calibration Certificate See chapter Calibration Certificate User Manual This document 1 5 Calibration Certificate Each calibration ki...

Страница 6: ...ns are defined for all calibration standards gauges and the torque wrench in the specific data sheets 3 Standard Definitions 3 1 Standard Definitions Standard definitions provide the data needed to ma...

Страница 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously...

Страница 8: ...setting parameters Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the d...

Страница 9: ...vailable In cases when the two VNA test ports have different connector systems installed e g RPC N 50 and RPC 3 50 UOSM can handle this situation A second calibration kit is needed Calibration Procedu...

Страница 10: ...calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following services are...

Страница 11: ...ood connector repeatability and its effect on the overall uncertainty of a measurement procedure should encourage careful inspection interface gauging and handling of coaxial connectors Prior to use a...

Страница 12: ...the utmost importance that connectors do not damage the test equipment interfaces to which they are connected Poor performance of many coaxial devices and cable assemblies can often be traced to poor...

Страница 13: ...tical limits should be done in different orientations averaging up to 4 single measurements When a gauge block is used to initially adjust the gauge a torque spanner should be used to tighten up the c...

Страница 14: ...ger com Tel 49 8684 18 0 Email info rosenberger com Page 14 17 Dieses Dokument ist urheberrechtlich gesch tzt This document is protected by copyright Rosenberger Hochfrequenztechnik GmbH Co KG Figure...

Страница 15: ...in which the centre conductor extends beyond the outer conductor mating plane This may result in buckling of the socket contacts or damage to the internal structure of a device due to the axial force...

Страница 16: ...l centre conductors are within the observed pin depth values in Table 2 2 Refer to 4 2 Interface Dimensions and Gaging 5 Carefully align the connectors The male connector centre pin must slip concentr...

Страница 17: ...en disconnecting a sliding load leave the centre conductor pullback handle in the locked position Use the combi wrench to prevent the device body from turning Use torque wrench to loosen the connectin...

Отзывы: