User Manual
RPC-N 50
Calibration Kit
05CK001-150 and 05CK010-150
05CK001-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
10 / 17
D
ie
se
s
D
ok
um
e
nt
is
t
u
rh
eb
e
rr
ec
h
tli
ch
g
es
ch
üt
zt
●
T
hi
s
d
o
cu
m
e
n
t i
s
p
ro
te
ct
ed
b
y
co
py
ri
g
ht
●
R
os
en
b
er
ge
r
H
o
ch
fr
eq
u
en
zt
ec
h
ni
k
G
m
bH
&
C
o
. K
G
5. Calibration Kit Service
5.1 Calibration interval
Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the
wear and the actual requirements a different calibration interval might be suitable. This lies in the
responsibility of the user.
5.2 Spare parts
Each part of this kit can be ordered separately as spare part with reference to the part number
shown in Table 1-1 / Table 1-2 and the quantity desired. Telephone or send your order to your
local Rosenberger sales office.
5.3 Where to send a Kit for Re-Calibration / Repair
In case of re-calibration or repair contact your local Rosenberger sales office
5.4 Extent of Rosenberger Re-Calibration Services
Every calibration kit sent for re-calibration or repair the following services are performed
Cleaning of connector interfaces
Optical check of connector mating planes, contact pins and contact fingers
Check of connector gauge for impermissible protrusion and significant recession
Electrical measurement of all calibration standards
Measurement of torque wrench torque
Measurement of gauges
Issuing of calibration certificates including measurement values and uncertainties
Individual optimizing of capacitive coefficients C0 to C3 for all Open calibration standards and
short inductance coefficients L0 to L3 for all Short calibration standards for best fitted model
based standard definitions.
In case of any defect or abnormality a cost estimation will be issued before continuing repair work.
If a calibration standard has to be replaced or modified, the incoming status is additionally reported
in the calibration certificate. This service is excluded when this measurement may damage the
Rosenberger measurement setup.