6.2.2 Procedure
1.
Remove the sensor from the process. Place it in an insulated container of water along with a calibrated ther-
mometer. Submerge at least the bottom two inches of the sensor. Stir continuously.
2.
Allow the sensor to reach thermal equilibrium. For some sensors, the time constant for a change in tempera-
ture is 5 min., so it may take as long as 30 min. for temperature equilibration.
3.
If the sensor cannot be removed from the process, measure the temperature of a flowing sample taken from
a point as close to the sensor as possible. Let the sample continuously overflow an insulated container hold-
ing a calibrated thermometer.
4.
Change the Solu Comp II display to match the calibrated thermometer using the procedure below.
MODEL SOLU COMP II
SECTION 6.0
CALIBRATION
a.
Press MENU. The main menu screen appears. Choose Calibrate.
b.
Choose Sensor1 or Sensor2. For a single input configuration, the
Sensor1 Sensor2 screen does not appear.
c.
Choose Temp.
d.
If the analyzer was programmed in Section 5.6 to use the actual process
temperature, the screen at left will appear. To calibrate the temperature,
change the number in the second line to match the temperature meas-
ured with the standard thermometer. Press ENTER. The display
changes to the screen in step f.
If the analyzer was programmed to use a temperature entered by the
user, the screen shown in step e will appear.
e.
If the analyzer was programmed in Section 5.6 to use the temperature
entered by the user, the screen at left will appear. Change the tempera-
ture to the desired value, then press ENTER. The analyzer will use the
temperature entered in this step in all measurements and calculations, no
matter what the true temperature is.
f.
The screen at left will appear. Press EXIT.
g.
Choose the other sensor and enter the temperature.
h.
To return to the main display, press MENU followed by EXIT.
CCaalliibbrraattee
Hold
Program
Display
CalSensor1?
Measurement
TTeemmpp
CalSensor1?
Measurement
TTeemmpp
Calibrate?
Sensor1
SSeennssoorr22
Calibrate?
SSeennssoorr11
Sensor2
Live
25.0
°
C
CalS1
++
25.0
°
C
ManualTemp?
S1: +25.0
°
C
30