Section
1-1
1
Introduction
OVERVIEW
This manual provides information on the operation, maintenance, and
calibration of the Rosemount Model P300 Pressure Validator. It is
organized into the following sections:
Section 1: Introduction
This section provides a manual overview, description of the Rosemount
Model P300 Pressure Validator, safety messages, and shipping list.
Section 2: Hardware Functions
This section describes power sources and battery installation and gives
a description of the pneumatic controls and the electrical control panel.
Section 3: Software Functions and Setup
This section describes the validator’s software functions, including
using the keypad, reading the display panel, setting parameters, and
setting the display.
Section 4: Validation Procedures
This section describes validation hookups for externally powered and
validator-powered procedures.
Section 5: Specifications and Reference Data
This section provides all validator specifications and data about the
validator’s internal configuration.
Section 6: Troubleshooting and Maintenance
This section explains the basic error messages and how to interpret
them. It also contains information on care and maintenance of the
validator.
Appendix A: Calibration
This section describes calibration procedures for the Rosemount Model
P300 Pressure Validator.
PRODUCT DESCRIPTION
The Model P300 Pressure Validator is a portable testing device
intended for use in remote locations for pressure testing and validation
of pressure transmitters and systems.
The Model P300 uses an internal pressure transducer to measure the
pneumatic pressure applied to the outlet port. Pressure is generated by
a hand-operated pump capable of supplying pressures up to 300 psi
gauge. An optional changeover valve generates negative pressures to
–12.75 psi gauge.
The electrical features of the validator include a fully isolated 24 V dc
power supply, a fully floating constant current source, a voltmeter and a
milliammeter. The 24 V dc power supply is designed to power external
devices at a load current of 25 mA.
Содержание P300
Страница 1: ...Model P300 Pressure Validator Installation and Operation Manual 00809 0100 4775 English Rev AA ...
Страница 2: ......
Страница 4: ...Rosemount Model P300 Pressure Validator ii ...
Страница 14: ...Rosemount Model P300 Pressure Validator 2 6 ...
Страница 20: ...Rosemount Model P300 Pressure Validator 3 6 ...
Страница 30: ...Rosemount Model P300 Pressure Validator 4 10 ...
Страница 34: ...Rosemount Model P300 Pressure Validator 5 4 ...
Страница 36: ...Rosemount Model P300 Pressure Validator 6 2 ...
Страница 45: ...I 3 Index ...
Страница 46: ...Rosemount Model P300 Pressure Validator I 4 ...
Страница 47: ... 1 ...