INSTRUMENT
SPECIFICATION
TABLE 5-1. Pressure Specification and
Ranges.
Pressure Accuracy
Combined nonlinearity, hysteresis, and repeatability
All pressure range versions ±0.075% Full Scale.
Pressure Temperature Effects
All versions 32 to 104 °F (0 to 40 °C) ±0.5%.
Pressure Output
Type: Handpump and Volume Adjuster.
Pump Range: –12.75 to 300 psi gage. (–0.88 to 20.7 bar).
Pressure/Vacuum Changeover: Manual Switch (Pressure/Vacuum
only).
Maximum Working Pressure
The pressure can be exceeded by up to two times its range, causing
negligible validation change.
Burst Pressure
Option Code 4 (150 inH
2
O): 184 psi (12.7 bar).
Option Code 5 (750 inH
2
O): 350 psi (24 bar).
Option Code 6 (100 psi): 375 psi (25.9 bar).
Option Code 7 (300 psi): 375 psi (25.9 bar).
ELECTRICAL
SPECIFICATIONS
Electrical Safety
Meets EN61010-1 as applicable.
Power Supplies
External dc supply: 3.5 to 12V (Max) dc.
Internal batteries: Four D cell or NiCad rechargeables.
Battery life
Standard D cell: 40 hours.
Alkaline D cell: 70 hours.
NiCad (4 Ah): 18 hours.
Battery charge current (NiCad only): 400 mA.
Voltage Measurement Range
No.
Range
4
150 in
5
750 in
6
100 psi
7
300 psi
Содержание P300
Страница 1: ...Model P300 Pressure Validator Installation and Operation Manual 00809 0100 4775 English Rev AA ...
Страница 2: ......
Страница 4: ...Rosemount Model P300 Pressure Validator ii ...
Страница 14: ...Rosemount Model P300 Pressure Validator 2 6 ...
Страница 20: ...Rosemount Model P300 Pressure Validator 3 6 ...
Страница 30: ...Rosemount Model P300 Pressure Validator 4 10 ...
Страница 34: ...Rosemount Model P300 Pressure Validator 5 4 ...
Страница 36: ...Rosemount Model P300 Pressure Validator 6 2 ...
Страница 45: ...I 3 Index ...
Страница 46: ...Rosemount Model P300 Pressure Validator I 4 ...
Страница 47: ... 1 ...