Figure 5-2. Fault 2, Shorted Thermocouple
b.
Fault 2, Shorted Thermocouple. The HEATER
T/C LED flashes twice, pauses for three seconds,
and repeats (Figure 5-2).
1. Using a multimeter, measure across TP3+
and TP4-.
2. If the reading is 0 ±0.5 mV, then a shorted
thermocouple is likely.
3. Remove power and disconnect J1.
4. Measure from TP3+ to TP4-. The reading
should be approximately 20 Kohms.
5. If so, the short is not on the PC board. See
paragraph 6-7, Heater Strut Replacement.