IB-106-340
7-8
7-8.
DEFINING A TIMED CALIBRATION VIA
HART. Use the following procedure to specify a
time interval (in hours) at which the Oxymitter 4000
will be automatically calibrated.
NOTE
To select a menu item, either use the up and
down arrow keys to scroll to the menu item
and press the right arrow key or use the
number keypad to select the menu item
number.
To return to a preceding menu, press the
left arrow key.
a.
From the DEVICE SETUP screen, select
DETAILED SETUP.
b.
From the DETAILED SETUP screen, select O2
CALIBRATION.
c.
From the O2 CALIBRATION screen, select
menu item 6, CAL MODE. Set the CAL MODE
to AUTO.
d.
Return to the O2 CALIBRATION screen and se-
lect menu item 7, CAL INTRVL.
e.
At the prompt, input a time interval (in hours) at
which an automatic calibration will occur and
press ENTER.