Figure 5-5. Fault 5, High High Heater Temp
e.
Fault 5, High High Heater Temp. The
HEATER LED flashes twice, pauses for three
seconds, and repeats (Figure 5-5).
1. The high high heater temp alarm will
activate when the thermocouple produces a
voltage of 37.1 mV (1652°F/900°C).
2. The triac and the temperature control may
be at fault.
3. Remove power. Allow Oxymitter 4000 to
cool for five minutes. Restore power.
4. If the condition repeats, replace the
electronic assembly per paragraph 6-5b,
Electronic Assembly Replacement.