If the laser pulse amplitude drops below the laser threshold (default 0.2 V), the Laser
Healthy alarm turns red, and the LASER bulb on the front of the control system is red. At
this point, the line stops, because of the red light; clean the mirrors.
Error Message for Pulse Amplitude below Threshold
Figure 7-2:
Procedure
1.
Switch off the BLOWER ISOLATOR and compressed air supply.
2.
Open the upper enclosure to see the red measurement cell jacket.
3.
Release the two clamps on top of the cell lid.
4.
Remove the top cover of the cell jacket.
5.
Remove the black cell cover.
6.
Use only Thor Labs lens cleaning tissue (P-5000-0613) to gently wipe the mirrors on
each side of the cell (see
Figure 7-3
).
CAUTION!
EQUIPMENT DAMAGE
Avoid skin contact with the mirrors as oils can be deposited.
Cleaning the Cell Mirrors
Figure 7-3:
Maintenance
34
Rosemount CT4215