MODEL 54e pH/ORP
SECTION 3.0
WIRING
To prepare the cable for sensor wiring:
1.
Use only the cable specified. Figure 3-4 shows
the 10 conductor cable with 3 shields
(9200273).
2.
Strip back the PVC jacket 4 in. or far enough
to access the eleven screw terminals in the
junction box, or the eleven terminals in the
controller. Separate the two coaxial cables and
prepare each as follows:
2a. Strip back the insulating black sheath
about 1½ in.
2b. Separate the braid from the inner black
conductive sheath.
2c. Solder an insulated wire to the braid.
2d. Strip the black conductive sheath 1 in. to
expose the colored (orange or gray) cable
inside.
2e. Insulate the exposed black sheath and braid area to prevent shorts.
3.
Strip ¼ in. of insulation on each conductor for terminal mounting. Insulate the exposed foil shields with heat
shrink sleeves or electrical tape. Be sure that heat shrink overlaps the exposed metal end where the foil
shield has been cut. Shields must not be shorted together. The sensor will not work if foil shields or drain
wires are not electrically isolated from each other.
FIGURE 3-4. Cable Dressing
FIGURE 3-5. Wiring for Sensors that have Solution Grounds (Models 381+, 385+, 396P, 396R)
INSIDE FRONT PANEL VIEW
11