MODEL 54e pH/ORP
SECTION 5.0
SOFTWARE CONFIGURATION
TABLE 5-1. pH Settings List (continued)
ITEM
RANGES
FACTORY SETTINGS
USER SETTINGS
5. Alarm 3 Control
(a) Activation Method
Process/Temp
Process
_______
(b) Alarm 3 Control Mode
Normal/TPC
Normal
_______
6a. Alarm 3 Setup (Normal)
(a) Configuration
Low alarm/High alarm/Off
High
_______
(b) Hysteresis (deadband)
0 - 5.00 pH
0.01 pH
_______
(e) Delay Time
0-99 sec
0 sec
_______
(d) Relay Fault
Open/Closed/None
None
_______
6b. Alarm 3 Setup (TPC)
(a) Setpoint
-2 to 16 pH or -15 to 130°C
7 pH
_______
(b) Proportional
0-299.9%
100.0%
_______
(c) Integral
0-2999 sec
0 sec
_______
(d) Derivative
0-299.9%
0.0%
_______
(e) Time Period
10-2999 sec
30 sec
_______
(f) LRV (100% On)
-2 to 16 pH or -15 to 130°C
2 pH
_______
(g) URV (100% Off)
-2 to 16 pH or -15 to 130°C
0 pH
_______
(h) Relay Fault
None/Open/Closed
None
_______
7. Alarm 4 Control
(a) Alarm
Fault/Off
Fault
_______
8. Feed Limit Timer
(a) Feed Limit
Disable/alarm 1/alarm 2/alarm 3
Disable
_______
(b) Timeout Value
0-10,800 sec
3600 sec
_______
C. Alarms (Section 5.7)
9. Interval Timer
(a) Timer (selection)
Disable/alarm 1/alarm 2/alarm 3
Disable
_______
(b) Timer (activation method)
Time activated/Imped. activated
Time activated
_______
(c) Interval
0-999.9 hr
24.0 hr
_______
(d) Repeats
1-60
1
_______
(e) On Time
0-2999 sec
120 sec
_______
(f) Off Time
0-2999 sec
1 sec
_______
(g) Recovery
0-999 sec
600 sec
_______
D. Diagnostics (Section 5.8)
1. Diagnostics (Glass and Reference)
On/Off
000 (no security)
_______
2. Glass Imp(edance) High Setpoint
0-2000 MOhms (0 disables)
000 (no security)
_______
3. Glass Imp(edance) Low Setpoint
0-900 MOhms (0 disables)
4. Ref(erence) Imp(edance) High
0-140 KOhms (0 disables)
000 (no security)
_______
5. Zero Offset
0-999 mV (0 disables)
000 (no security)
_______
6. Cal(ibration) Warn(ing)
0-500% (0 disables)
7. Imped(ance Temperature) Comp(ensation)
On/Off
000 (no security)
_______
E. Auto Calibration (Section 5.9)
1. Autocal (Buffer List Selection)
Manual/Standard/DIN 19267
000 (no security)
_______
2. Stabilize pH (Auto Calibration)
.01-.50 pH
000 (no security)
_______
3. Stabilization Time (Auto Calibration)
0-30 sec
000 (no security)
_______
F. Security (Section 7.1)
1. Lock all
000-999
000 (no security)
_______
2. Lock Program (Lock all except Calibrate)
000-999
000 (no security)
_______
3. Lock Config. (Lock all except Calibrate,
Output setpoints (PID), Simulated Tests
Alarm Setpoints, and Rerange Outputs)
000-999
000 (no security)
_______
D. Solution Temperature Compensation (Section 7.3)
1. Temp(erature) Coeff(icient)
-0.044 to 0.028pH/°C
000 (no security)
_______
2. Operate Iso(potential pH)
-1.35 to 20.12 pH
000 (no security)
_______
3. Sensor Iso(potential)
0-14 pH
000 (no security)
_______
By changing the standard output configuration, you can set up the Model 54e pH/ORP to perform a wide variety of control and monitoring tasks.
The configuration procedures allow you to program the controller to meet the specific control and monitoring requirements of your particular
plant. This is done by recording the desired configuration parameters on the List of Settings Form and then actually configuring them by using
the keys on the controller front panel.