42
MODEL 1054B pH/ORP
SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
8.2 TROUBLESHOOTING.
The Model 1054B
Analyzer is designed with state-of-the-art micro-
processor circuitry, making troubleshooting simple
and direct. Subassembly replacement, i.e. printed
circuit board replacement, is all that is usually
required.
8.2.1 Installation Failure.
If failure does occur, com-
plete the following steps:
1.
Check for a fault flag. If a fault condition exists,
refer to
for the fault mnemonic explana-
tion.
2.
Check wiring connections for proper installation.
3.
Refer to Troubleshooting
(ORP). The table is arranged with the
most common problems listed first.
8.2.2 LCD/LED Display Test.
Display Mnemonic
dtS
.
Selecting this option will activate all the display seg-
ments. This item is used if a faulty display is suspect-
ed. Refer to
1.
Enter Set Mode by pressing
ACCESS
key twice.
2.
SCROLL
(
é
) until
dtS
appears on the display.
3.
SELECT.
8.2.3 Software Version.
Display Mnemonic
UEr
.
Selection of this item will display the software revision
level of the CPU. This number may be requested by
factor y service personnel if troubleshooting is
required.
1.
Enter Set Mode by pressing
ACCESS
key twice.
2.
SCROLL
(
é
) until
UEr
appears on the display.
3.
SELECT.
(i.e., 2.0.6 displayed.)
8.2.4 Sensor Troubleshooting.
In addition to the
fault mnemonics that relate to a possible sensor
problem (
display the input in millivolts generated from the sen-
sor. See Table 8-5 for how the millivolt input relates to
pH. If the analyzer values do not agree within about
40 to 60 mV of those in the table, the analyzer has
been incorrectly standardized for pH, the reference
may be significantly poisoned, or some other signifi-
cant problem exists. To read the millivolt input, go to
Section 8.2.5.
For an additional diagnostics tool you can look at the
rES
impedance value as described in
8.2.5 Electrode Input (1054B pH).
Display Mnemonic
Ein
. When selected, the analyzer displays the millivolt
input from the sensor. The displayed value is not tem-
perature corrected.
1.
Press the
ACCESS
key twice.
SEt
will briefly dis-
play followed by
Ein
. Not required if already in
set menu.
2.
SELECT
this item to read the mV input.
3.
Press the
PV
key to return to normal operation.
8.2.6 CPU Board Replacement.
If there is a problem
with the CPU board resulting in its replacement, spe-
cific procedures (included with the order) for calibrat-
ing the new board must be followed exactly or the
microprocessor will be improperly programmed.
Should this occur, it will be necessary to return the
analyzer to the factory for reprogramming.
8.2.7 Power Board Replacement.
If it becomes nec-
essary to replace the power board, the CPU board
will need to be recalibrated following specific proce-
dures that are included with the power board. Failure
to follow these procedures exactly will cause the
microprocessor to be improperly programmed and
require the return of the analyzer to the factory for
reprogramming.
8.3 INSTRUMENT MAINTENANCE.
To maintain the
appearance and extend the life of the enclosure, it
should be cleaned on a regular basis using a mild
soap and water solution followed by a clean water
rinse.
Input to Analyzer
pH
in Millivolts
15°C
(59°F)
25°C
(77°F)
50°C
(122°F)
80°C
(176°F)
0
396
410
444
486
1
340
351
381
416
2
283
293
317
347
3
226
234
254
277
4
170
176
190
208
5
113.2
117.1
127.0
138.7
6
56.6
58.6
63.5
69.4
7
0
0
0
0
8
– 56.6
– 58.6
– 63.5
– 69.4
9
– 113.2
– 117.1
– 127.0
– 138.7
10
– 170
– 176
– 190
– 208
11
– 226
– 234
– 254
– 277
12
– 283
– 293
– 317
– 347
13
– 340
– 351
– 381
– 416
14
– 396
– 410
– 444
– 486
TABLE 8-5
Sensor Input to Analyzer Verses
pH at Four Temperatures*
* These values do not apply for non-glass electrodes.
Содержание 1054B
Страница 3: ......
Страница 11: ...6 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 2 Panel Mounting Tab Installation DWG NO REV 41054A26 A ...
Страница 15: ...10 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 6 Electrical Wiring DWG NO REV 41054B03 C ...
Страница 16: ...11 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 7 Integral Preamp Wiring DWG NO REV 41054B32 B ...
Страница 52: ...FOLD ALONG DOTTED LINES ...