14
MODEL 1054B pH/ORP
SECTION 3.0
DESCRIPTION OF CONTROLS
SECTION 3.0
DESCRIPTION OF CONTROLS
3.1 KEYBOARD FUNCTIONS.
All functions of the 1054B are accessed through key-
board entry routines. The analyzer uses no switches or
potentiometers.
The four keys across the top row and the
CAL
(pH
only) and
ENTER
keys are dual function. One press of
the key will display the value of the function shown on
the lower portion of the key. A quick double press of
the key will display the value of the function shown on
the upper portion of the key. Each of these keys have
read functions that can be accessed without security
code entry. Each key also has a calibration or set func-
tion when used with the
SELECT
key. This function
requires entry of the security code when the security
feature is active. (Refer to
security.)
NOTE
When no key is pressed for a period of 60
seconds the analyzer will default to read-
ing pH.
CAUTION
The
HOLD
function and the
CAL
func-
tion are not read functions. Refer to
of the pH/ORP sensor is achieved by
pressing the
key.
Std
displays to acknowledge the
standardize function, followed by the
Numeric Display for user input. Entering the known
pH/ORP of the measured solution will cause the ana-
lyzer to restandardize the sensor. The pH glass slope
value will not be changed. Refer to
B. Standardize Temperature.
Stand-
ardization of the temperature is achieved
by pressing the
TEMP
key once, followed
immediately by pressing the
SELECT
key.
AdJ
displays to acknowledge the
standardization function, followed by the
Numeric Display for user input. Entering the known
temperature of the measured solution will cause the
analyzer to restandardize the temperature reading.
Refer to
C. Alarm 1 and Alarm 2 Setpoint.
The
alarm setpoint may be adjusted by
pressing the
ALARM 1
or
ALARM 2
key
once, followed by pressing the
SELECT
key.
AdJ
displays, followed by the
Numeric Display for user input. Refer to
D. Current Output – Zero Setpoint.
The
zero point (0 or 4 mA) of the pH output
range is adjusted by pressing the
ZERO
key twice, followed by pressing the
SELECT
key.
AdJ
displays, followed by
the Numeric Display for user input. Refer
E. Current Output – F.S. Setpoint.
The
full scale point (20 mA) of the pH out-
put range is adjusted by pressing the
key twice, followed by pressing
SELECT
key.
AdJ
displays, followed
by the Numeric Display for user input. Refer to
F. Two Buffer Calibration (pH).
buffer calibration is initiated by pressing
the
G. pH Glass Slope.
The millivolt output
per pH unit is adjusted by pressing the
CAL
key twice, followed by pressing the
SELECT
key once.
AdJ
displays, fol-
lowed by the Numeric Display for user
input. Refer to
H. Simulate Current Output.
The pH
output can be simulated by pressing
the
PV
key twice, followed by pressing
the
SELECT
key. The Numeric Display
appears for user input. Refer to
ZERO
ALARM 1
HOLD
TEMP
CAL
ZERO
ALARM 1
OUTPUT
PV
Содержание 1054B
Страница 3: ......
Страница 11: ...6 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 2 Panel Mounting Tab Installation DWG NO REV 41054A26 A ...
Страница 15: ...10 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 6 Electrical Wiring DWG NO REV 41054B03 C ...
Страница 16: ...11 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 7 Integral Preamp Wiring DWG NO REV 41054B32 B ...
Страница 52: ...FOLD ALONG DOTTED LINES ...