MODEL 1054B pH/ORP
SECTION 5.0
START-UP AND CALIBRATION
5.8. ORP STANDARDIZATION (1054B ORP only).
For maximum accuracy, the sensor should be stan-
dardized in the ORP standard process or process
grab sample.
Step 1:
Ensure that the Model 1054B is in hold (Hold
flag showing). If not, Press the H
HO
OL
LD
D key twice.
Step 2:
Using a ORP standard, determine the ORP
of the solution.
Step 3:
Ensure that the sensor’s measurement por-
tion is in the process or grab sample.
Step 4:
Allow the reading to stabilize.
Step 5:
Press the
PV
or
CAL
key once, then Press
the
SELECT
key.
Std
will show followed by the
Numeric Display with digit flashing.
Step 6: SCROLL
(
é
)
and
SHIFT
(
ç
)
to display the
proper ORP value then
ENTER
it into memory.
Step 7:
Press the
HOLD
key twice to remove the
Model 1054B from hold. The 1054B ORP is now
ready for operation.
5.9. SENSOR MAINTENANCE.
Before preforming
maintenance or repair of the sensor, the Model
1054B should be placed in hold. This will place the
current output and relays in the state set in the
default setup. Press the
HOLD
key twice. The hold
flag will show to indicate the hold condition. Always
calibrate (
) after cleaning or repair of the
pH sensor.
Содержание 1054B
Страница 3: ......
Страница 11: ...6 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 2 Panel Mounting Tab Installation DWG NO REV 41054A26 A ...
Страница 15: ...10 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 6 Electrical Wiring DWG NO REV 41054B03 C ...
Страница 16: ...11 MODEL 1054B pH ORP SECTION 2 0 INSTALLATION FIGURE 2 7 Integral Preamp Wiring DWG NO REV 41054B32 B ...
Страница 52: ...FOLD ALONG DOTTED LINES ...