4. Place the patient tubing to the appropriate position of patient and start to 

suck sputum aspirate.

5. When the sputum reaches the safe full level of collection (800 cc), the 

suctioning will stop sucking automatically (float shut off).

6. Switch the power switch to “O” position (Off), shut off the suction motor.
7. Remove the sputum from collection bottle after the motor stops turning 

completely. (Refer to the “Cleaning” section below for the instructions on how 
to clean the collection bottle.)

8. Turn on the unit and start to run the suction unit again.

 NOTE

1. If the unit sustains a severe drop, validity of the gauge 

must be checked. Inspect if the float shut-off is still 
functioning properly when the liquid in the collection 
reaches the safe full level.

2. Always transport unit with the vacuum adjust knob 

rotated fully clockwise in case unit is dropped.

 

 CAUTION

1. When automatic float shut-off is activated, contents of the 

collection bottle should be emptied. Further suctioning 
could cause damage to the vacuum pump.

2. Should fluid be aspirated back into the unit, discontinue 

use immediately and send your aspirator to an authorized 
service center for a detailed inspection as the vacuum 
pump may have been damaged.

CLEANING

CLEAN COLLECTION BOTTLE

1. Turn off the aspirator and allow vacuum to drop, then disconnect AC power 

source.

2. With the collection bottle still in the holder, remove the collection bottle lid. 

The bottle can now be taken out of the holder to be emptied.

   NOTE

Collection bottle should be emptied after each use.

3. The collection bottle should be thoroughly cleaned after each use by one of 

the following three methods:

  a. 

Wash in a hot water/dishwashing detergent solution and rinse with  

 

 

clean, hot tap water. Then wash in one part vinegar to three parts hot    

 

water solution. Rinse with hot tap water and air dry.

  b. 

Wash with rubbing alcohol and air dry.

      WARNING

If the device has been operating continuously for 30 
minutes or more, turn off power and allow the device to 
cool for at least 30 minutes before restarting

.

 

Содержание ASP-ROS

Страница 1: ...ROSCOE ASP ROS HEAVY DUTY ASPIRATOR Manual MODEL ASP ROS Read this manual before operating the aspirator Save these instructions for future reference...

Страница 2: ...cations Classifications 10 10 Troubleshooting 11 11 Warranty 12 CAUTION Federal law restricts this device to sale by or on order of a physician or any other practitioner licensed by the law of the Sta...

Страница 3: ...THIS MANUAL FOR FUTURE REFERENCE DANGER This medical device is not intended to be used as a life sustaining supporting device DANGER To reduce the risk of electrocution Do not use while bathing Do no...

Страница 4: ...2 AC POWER SOURCE CAUTION Item ASP ROS is used with a power supply allowing opera tion on AC voltage AC 110 120 V 60 Hz The correct power source must be used to connect to adaptable wall power 3 ELECT...

Страница 5: ...ION Your Heavy Duty Suction Pump is a professional medical suctioning device that produces a maximum vacuum of 675 mmHg The unit is equipped with an 800cc collection canister and has been designed for...

Страница 6: ...Long intermediate tubing 3 Inlet 4 Power switch 5 Short intermediate tubing 6 Collection container lid 7 Collection container holder 8 Suction tubing 9 Collection container 10 Vacuum port 11 Patient p...

Страница 7: ...eck the specification label on the bottom of the suction unit to ensure that the voltage and current indicated on the unit correspond with the voltage and current available Make sure the power switch...

Страница 8: ...the collection bottle should be emptied Further suctioning could cause damage to the vacuum pump 2 Should fluid be aspirated back into the unit discontinue use immediately and send your aspirator to...

Страница 9: ...3 Rinse with hot tap water and air dry 4 Keep the outer surface of the tubing clean by wiping with a clean damp cloth WARNING The waste in collection bottle may cause respiratory infection which must...

Страница 10: ...for the filter Improper substitution may lead to contamination or poor performance COLLECTION CANISTER MAINTENANCE 1 Discard contents at least daily 2 Wash canister with hot water and a mild dish det...

Страница 11: ...e 4 Allow machine to cool for 30 minutes before resuming operation 5 If the fibrous material inside the inline filter is wet stained or damaged replace the filter immediately Motor runs no vacuum 1 Ve...

Страница 12: ...repair or replace parts at issue provided the claim is bona fide It shall be the responsibility of the purchaser to pack and return the equipment in a manner to avoid shipping damage The shipment must...

Страница 13: ...ROSCOE ASP HD ASPIRADOR PARA TRABAJO PESADO Manual MODELO ASP HD Lea este manual antes de usar el aspirador Guarde estas instrucciones para una referencia futura...

Страница 14: ...Instrucciones de Limpieza 28 29 8 Mantenimiento 30 9 Especificaciones Clasificaciones 30 10 Resoluci n de Problemas 30 11 Garant a 30 NOTA Revise TODAS las partes del equipo una vez despachado para ve...

Страница 15: ...ca una situaci n potencialmente peligrosa que si no se evita puede causar lesiones leves o moderadas al usuario o al paciente o da os al dispositivo u otra propiedad NOTA Informaci n a la que debe pre...

Страница 16: ...DESCARGAS ELECTRICAS SI SE REALIZAN INCORRECTAMENTE SOLAMENTE UN TECNICO CALIFICADO DEBE REALIZAR EL SERVICIO DE MANTENIMIENTO Y REPARACI N DE ESTE SUCCIONADOR DE GLERAS PELIGRO El succionador de gle...

Страница 17: ...a evitar da os f sicos es necesario que el equipo funcione adecuadamente El uso inadecuado de este equipo puede causar lesiones personales y o da o de tejidos Este equipo no debe ser utilizado a menos...

Страница 18: ...llevar y lavar El seguimiento de los procedimientos recomendados en este manual para el funcionamiento y el mantenimiento de este equipo maximizar la vida til de este producto Referirse a los diagrama...

Страница 19: ...Bot n de ajuste del vac o 3 Fusibles 4 Cable de alimentacion 5 Interruptor de encendido 6 Cojines 7 Filtro 8 Cobertura para Conexi n de tuber a 9 Paciente Conexi n de tuber a 10 C nula para el pacien...

Страница 20: ...e disponibles Aseg rese que el interruptor de encendido ubicado en el lateral del equipo est en OFF antes de conectar el equipo a la fuente de poder 2 Coloque el interruptor en la posici n I encendido...

Страница 21: ...reloj en caso que el equipo se caiga PRECAUCI N 1 Cuando el sensor de sobre flujo est activado los contenidos que se encuentran en los vasos recolectores deben ser vaciados Sobre succi n puede causar...

Страница 22: ...ber a del equipo 2 La tuber a debe ser enjuagada minuciosamente con agua caliente seguido de una soluci n de 1 parte de vinagre para 3 partes de agua despu s de cada uso 3 Enjuague con agua caliente y...

Страница 23: ...arriba Los filtros adicionales se puede comprar en una tienda distribuidora autorizada ESPECIFICACIONES CLASIFICACIONES Dimensiones L 375 x W 175 x H 250 mm 14 8 x 6 8 x 9 8 in Peso 5 5 kg 12 1 lb Asp...

Страница 24: ...regladas por el usuario Si usted cree que su equipo no est funcionando adecuadamente por favor t mese unos minutos para revisar las siguientes causas ANTES DE DEVOLVERLO A LA TIENDA DISTRIBUIDORA DOND...

Страница 25: ...u garant a por favor contacte a un distribuidor autorizado por Compass Health Brands comerciante SERVICE AND REPAIR Su producto de la marca Roscoe est garantizado de ser libre de defectos en materiale...

Страница 26: ..._____________________________________________________________________ _______________________________________________________________________ __________________________________________________________...

Страница 27: ...Manufactured for Fabricado por Compass Health Brands Corp Middleburg Heights OH 44130 www compasshealthbrands com Phone Telefono 800 376 7263 Made in Taiwan Echo en Taiwan 42 ASP ROS_00 2020...

Отзывы: