P. Electromagnetic compatibility information
MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
For EQUIPMENT and SYSTEMS that are not LIFE -SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the
Romed Automatic Digital Blood Pressure Monitor
The Romed Automatic Digital Blood Pressure Monitor is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Romed Automatic Digital
Blood Pressure Monitor can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Romed Automatic
Digital Blood Pressure Monitor as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
The instructions for use are available
on our website in different languages.
18
-5 ºC
3 5 ºC