9
Smart Check by ROHO Operation Manual
How to Perform a Check:
1. You should be in your usual sitting position.
2. Press the Check Button.
3. Refer to the table below.
•
Perform a Check frequently, at least once a day!
•
If a cover was used while setting up the cushion, the same cover must be used
during Check.
•
Perform a Check when there is a significant change in altitude.
Check Instructions
Perform a Check after completing Setup (Smart Setup or Optional Setup).
Check Table
NOTE:
After you perform a Check, you may choose to perform a Hand Check to become familiar with your inflation range. See the
h
and
c
heck
G
uIde
.
NOTE:
You may be able to add or remove a small amount of air and still be at your cushion inflation setting. Smart Check saves your cushion
inflation setting with some room for adjustment.
Light
Status
Action
The cushion is within the saved
inflation setting.
No further action is required.
Light
Possible Causes
Action
Your weight may be unevenly
distributed.
Perform a weight shift to circulate air in the cushion, then press
the Check Button to get a new reading.
The quick disconnect is not securely
attached to the cushion.
Make sure the connection is secure, then press the Check Button
to get a new reading. See
t
roubleShootInG
in this manual.
The cushion needs more air.
Open the valve and add more air to the cushion (6 to 10 pumps).
Close the valve. Press the Check Button to get a new reading.
Repeat as needed.
If you are unable to make adjustments on your own,
discontinue use until assistance is available.
The cushion or the quick disconnect
on the cushion has a slow leak.
See TROUBLESHOOTING in the
ROHO
®
Single Compartment
Cushion with Sensor Ready
™
Technology Operation Manual
or
contact Customer Care.
Light
Cause
Action
The cushion has too much air.
Open the valve and remove air from the cushion until you see the
green light. Close the valve.
If you are unable to perform this task, seek assistance.
Symbols and Lights for Check
Red Light
“Remove Air.
See the Check
Table.”
Green Light
“Within the saved
inflation setting.”
Setup
Button
Check
Button
Red Light
“Action required.
See the Check
Table.”
DRAFT