AGILITY CARBON Back System Operation Manual
Cleaning and Disinfecting
Warning:
-
Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before use by a different individual: clean,
disinfect, and check product for proper functioning.
-
Follow safety guidelines on the chemical containers.
Cautions:
-
DO NOT use abrasives such as steel wool or scouring pads on components of AGILITY CARBON.
-
DO NOT use cleaning products containing petroleum or organic solvents [i.e. acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry
cleaning fluids, adhesive removers] on components of AGILITY CARBON.
= Wipe with damp cloth
AGILITY Component
To Clean:
To Disinfect:
Back shell and hardware:
Hand wash in warm (40°C) water
with soap and a sponge. Rinse
with clean water. Wipe dry with a
clean cloth and then allow to air
dry completely.
Wipe gently with a cloth
dampened with a household
disinfectant. Rinse with
clean water. Wipe dry with a clean
cloth and then allow to air dry completely.
Cover:
Notes: Remove the foam and
air insert from the cover prior to
cleaning and disinfecting.
Secure the hook and loop fastener
strips at both ends of the cover.
Turn the cover inside out.
Zip the cover closed to avoid snags.
Machine wash in warm (40°C)
water with mild detergent.
Tumble dry low.
Hand wash using quaternary
ammonium disinfectant (Follow
product label instructions).
Rinse thoroughly with clean
water. The cover may
also be machine washed in
hot (60°C) water to disinfect.
Tumble dry low.
Note: The method of disinfection should be
determined based on disinfection needs.
Foam pad, foam insert and
lumbar support pad:
Note: Remove the foam
components from the cover
prior to cleaning.
Wipe lightly with a damp cloth.
DO NOT use soap. DO NOT
submerge in water. Wipe off
with a clean cloth. Completely
air dry before replacing in cover.
Note: Disinfectants are NOT effective on porous
surfaces such as foam. Replace soiled or
contaminated foam components with authorized
ROHO foam components only. Replacement
foam components may be purchased separately.
Soiled or contaminated foam components MUST
NOT be used by different individuals.
Air insert:
Notes: Remove the air insert from
the cover and foam pad prior to
cleaning and disinfecting.
Make sure the valve is closed.
DO NOT allow water or cleaning
solution to enter the air insert.
Place in a large sink or basin
with warm (40°C) water.
Hand wash, using mild liquid
hand soap, hand dishwashing
liquid, laundry detergent, or a
multipurpose detergent (Follow product label
instructions). Gently scrub all cushion surfaces,
using a soft, plastic bristle brush, a sponge, or
a cloth. Rinse with clean water. Air dry. DO NOT
machine wash or dry. DO NOT expose to direct
sunlight.
Hand wash in warm
(40°C) water using
1 part household
liquid bleach per 9 parts
water. Rinse thoroughly
with clean water. Air dry.
DO NOT machine wash or dry.
DO NOT autoclave.
or
1:9
+