R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
439
Scale the diagram between -40 dB and +10 dB.
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:PDIVision <divisions>[,
'<trace_name>']
Sets the value between two grid graticules (value per division) for the diagram area <Wnd>. When a new
PDIVision value is entered, the current RLEVel is kept the same, while the top and bottom scaling is
adjusted for the new PDIVision value.
<Wnd>
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON
). This suffix is ignored if the optional
<trace_name> parameter is used.
<WndTr>
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED
. This suffix is ignored if the
optional <trace_name> parameter is used.
<divisions>
Value and unit for the vertical diagram divisions.
Range [def. unit]
Range and unit depend on the measured quantity, see
Units for DISPlay...
Commands
.
*RST value
Depending on the measured quantity. The default reference level for an S-
parameter displayed in a
dB Mag
diagram is 10 dB.
'<trace_name>'
Optional string parameter for the trace name, e.g. 'Trc4'. If this optional
parameter is present, both numeric suffixes are ignored (trace names must be
unique across different channels and windows).
*RST value
–
SCPI, Command
Types
Confirmed, with query.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11
'
Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the trace number 9
to it.
DISP:WIND2:TRAC9:Y:PDIV 5
or:
DISP:WIND2:TRAC:Y:PDIV 5, 'CH4TR1'
Set the value per division to 5 dB.
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:RLEVel <level>[,
'<trace_name>']
Sets the reference level (or reference value) for a particular displayed trace. Setting a new reference level