R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
393
Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11.
DISP:WIND:STAT ON
Create diagram area no. 1.
DISP:WIND:TRAC:FEED 'CH4TR1'
Display the generated trace in diagram area no. 1.
CALCulate<Ch>:PARameter:DELete '<string>'
Deletes a trace with a specified trace name and channel.
<Ch>
Channel number.
'<string>'
Trace name, e.g. 'Trc4'. See
Rules for trace names
in the
Trace Manager
description.
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11.
CALC4:PAR:CAT?
Query the traces assigned to channel 4. If
Ch4Tr1
is the only trace assigned to
channel 4, the response is 'CH4TR1,S11'.
CALC4:PAR:DEL 'CH4TR1'
Delete the created trace.
CALCulate<Ch>:PARameter:MEASure '<string>', 'S11' | 'Y-S11' | ... | 'Z-S11'
Assigns a measurement parameter to an
existing
trace.
Note:
To create a new trace and at the same time assign the attributes, use
CALCulate<Ch>:PARameter:SDEFine
. To display the trace, create a window
(
DISPlay:WINDow<Wnd>[:STATe] ON
) and assign the trace to this window
(
DISPlay:WINDow<Wnd>:TRACe:FEED
); see example below.
Traces must be selected to become active traces; see
CALCulate:PARameter:SELect
.
CALCulate<Ch>:PARameter:CATalog?
returns a list of all defined traces. You can open the trace
manager (
DISPlay:MENU:KEY:EXECute 'Trace Manager'
) to obtain an overview of all channels
and traces, including the traces that are not displayed.
<Ch>
Channel number of an existing channel containing the referenced trace.
'<string>'
Trace name, e.g. 'Trc4'. See
Rules for trace names
in the
Trace Manager
description.