R&S ZVL
GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
149
Grid type: Keep
Sweep time
Time domain resolution
Unambiguous range
Stop freq. and no. of
points
Freq. gap and no. of
points
Stop freq. and
approx. freq. gap
Remote
control:
CALCulate<Chn>:TRANsform:TIME:LPASs KFSTop |
KDFRequency | KSDFrequency
Time Gate
Switches the time gate defined via
Define Time Gate
on or off. An active time gate acts on the trace in
time domain as well as in frequency domain representation.
Gat
is displayed in the trace list while the time
gate is active.
The time gate is independent of the frequency window used to filter the trace prior to the time domain
transformation.
In time domain representation, you can use the time gate
settings in order to eliminate unwanted
responses in your signal. After switching back to the frequency domain, you will receive the frequency
response of your DUT without the contribution of the unwanted responses.
Remote control:
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe <Boolean>
Define Time Gate
The
Define Time Gate
defines the properties of the time gate used to eliminate unwanted responses that