R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
353
irrespective of the selected trace format).
<Trace_Hist_Count>
Number of sweep to be read. 1 denotes the last sweep acquired, 2
denotes the second-last and so forth.
Range [def. unit]
1 to sweep count defined via
[SENSe<Ch>:]SWEep:COUNt
[–]
*RST value
–
SCPI, Command Types
Device-specific, query only
Example (see also
Sweep History):
SWE:COUN 10
Define the number of sweeps (10) to be measured in single sweep
mode.
INIT:CONT OFF; INIT
Activate single sweep mode and start a single sweep in channel no. 1.
CALC:DATA:NSW? SDAT,5
Query the results of the 5
th
sweep.
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
<Chn>
Channel number used to identify the active trace.
Parameters
–
Range [def. unit]
The data is transferred in the data format defined via FORMat[:DATA]. The unit
is the default unit of the sweep variable (Hz or dBm or s)
.
*RST value
–
SCPI, Command
Types
Device-specific, query only
Example:
See CALCulate<Chn>:DATA?
CALCulate<Chn>:DLINe...
This subsystem controls the horizontal line used to mark and retrieve response values (display line).
CALCulate<Chn>:DLINe <numeric_value>
Defines the position (response value) of the horizontal line.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric suffix
is set to 1.
Parameters
See list of parameters below.
Range [def.
unit]
The data is transferred in the data format defined via FORMat[:DATA]. The unit is
the default unit of the measured parameter; see
CALCulate<Ch>:PARameter:SDEFine
.