R&S
ZVAX24
Performance Test Report
1312.7979.82 1.7
E-2
Option R&S R&S ZVAX-B211 (Combiner): Checking the reflection loss of
PORT 3
Only with Option R&S ZVAX-B211
Test equipment: - VNA (item 1)
- Test cables (item 2)
Test setup:
Connect VNA test port 1 to Port 3 SOURCE OUT
VNA settings:
- [
Preset
]
- [
Cal
: Cal Manager : (
Apply cal 3 from cal pool
) ]
- [
Mode
: R&S ZVAX Path Config : Combiner ]
- [
Pwr BW AVG
: Meas Bandwidth : 1 kHz ]
- [
Meas
: S21 ]
- [
Marker
: Marker 1 ]
- [
Marker
Funct
: Max Search ]
- [
Marker
: Marker 2 to Marker 8 : Marker Frequency : (
see Test Report
) ]
Test points:
10 MHz to 8 GHz, 8 GHz to 24 GHz, 201 points
Measurement:
Marker 1 : Read marker frequency
Marker 1 to Marker 8 : Read return loss S11 ( positive values)
Limit:
See Performance Test Report
Uncertainty:
See Performance Test Report
Option R&S ZVAX-B251 (Harmonic Filter Source Port 1): Checking the
transmission loss
Only with Option R&S ZVAX-B251
Test equipment: - VNA (item 1)
- Test cables (item 2)
- Only with option R&S ZVAX-B291: Adapter 3.5 mm f-f (item 4)
- Only with option R&S ZVAX-B291: Terminations 3.5 mm m (item 5)
Test setup:
Connect VNA test port 1 to Port 1 SOURCE IN and test port 2 to
Port 1 SOURCE OUT, if option R&S ZVAX-B291 is not installed
R&S ZVAX24 Test Port 1, if option R&S ZVAX-B291 is installed
VNA settings:
- [
Preset
]
- [
Cal
: Cal Manager : (
Apply cal 1 from cal pool
) ]
- [
Mode
: R&S ZVAX Path Config : Src 1 Harmonic Filter ]
- [
Start
: 1 GHz (8 GHz)]
- [
Stop
: 8 GHz (24 GHz)]
- [
Pwr BW AVG
: Meas Bandwidth : 1 kHz ]
- [
Meas
: S21 ]
- [
Marker
: Marker 1 ]
- [
Marker
Funct
: Max Search ]
- [
Marker
: Marker 2 to Marker 8 : Marker Frequency : (
see Test Report
) ]
Test points:
1 GHz to 8 GHz, 8 GHz to 24 GHz, 201 points
Measurement:
Marker 1 : Read marker frequency
Marker 1 to Marker 8 : Read transmission loss S21
Limit:
See Performance Test Report
Uncertainty:
See Performance Test Report
Содержание ZVAX24
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 70: ...Contents R S ZVAX24 1312 7979 82 I 2 2 E 1 ...
Страница 72: ......
Страница 74: ......
Страница 106: ...Contents R S ZVAX24 1312 7979 82 I 4 2 E 1 ...
Страница 110: ......
Страница 112: ......
Страница 114: ......
Страница 115: ...Spare Parts List ...
Страница 116: ......
Страница 117: ......
Страница 118: ......
Страница 119: ......
Страница 120: ......
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ...Mechanical Drawings ...
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......
Страница 139: ......
Страница 140: ...Circuit Diagram ...
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......