7.3.9 INITiate Commands
The
INITiate...
commands control the initiation of the trigger system and define
the scope of the triggered measurement.
................................................................................................ 908
...............................................................................................908
INITiate<Ch>[:IMMediate][:DUMMy]
.................................................................................909
INITiate<Ch>[:IMMediate]:SCOPe
................................................................................... 909
............................................................................................. 909
INITiate[:IMMediate]:ALL
Starts a new single sweep sequence in all channels. This command is available in sin-
gle sweep mode only (
OFF
). The data of the last sweep
(or of previous sweeps, see
Retrieving the Results of Previous Sweeps
) can be read
CALCulate<Chn>:DATA:NSWeep:FIRSt?
SDATa, <count>
.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlap-
Chapter 6.4, "Command Processing"
Example:
See
Usage:
Event
Manual operation:
See
INITiate:CONTinuous:ALL
<Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Parameters:
<Boolean>
ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
See
Manual operation:
See
"All Channels Continuous / All Channels on Hold"