GUI Reference
R&S
®
ZNB/ZNBT
255
User Manual 1173.9163.02 ─ 38
Sweeps
Selects the number of sweeps to be measured in "Single" sweep mode: the measure-
ment is stopped after the specified number of sweeps. This setting applies to the active
channel.
Tip:
If a
other than DEFAULT is used and "Sweep All Channels" is
selected in the
, then "Sweeps" applies to all channels.
Remote command:
[SENSe<Ch>:]SWEep:COUNt
[SENSe:]SWEep:COUNt:ALL
Restart Sweep
Stops the current measurement and restarts a measurement cycle. In "Single" sweep
mode a new single sweep sequence is started.
Remote command:
INITiate<Ch>[:IMMediate][:DUMMy]
INITiate[:IMMediate]:ALL
Restart Manager...
other than DEFAULT is selected (i.e. if some compatibility mode
All Channels Continuous / All Channels on Hold
Selects a common sweep control mode for all channels of the active recall set.
●
"All Channels Continuous": The R&S
ZNB/ZNBT continuously repeats the sweeps
in all channels.
●
"All Channels on Hold": The R&S
ZNB/ZNBT performs single sweeps, according to
the channel-specific number of "Sweeps".
These actions are only available for the DEFAULT
). For other
remote languages (compatibility modes) you can use the
instead.
Remote command:
Sweep Controller
Activates/deactivates the (resizable) "Sweep Info" dialog, which displays the current
sweep stage. This is particularly useful for long duration sweeps, executed in single
sweep mode: by observing the dialog it is easy to realize when the sweep is done.
Sweep Softtool