GUI Reference
R&S
®
ZNB/ZNBT
618
User Manual 1173.9163.02 ─ 38
5.19.3.1
Controls on the Remote Settings Tab
IP Address
Displays the current IP address of the R&S
ZNB/ZNBT. By default, the analyzer is con-
figured to use dynamic TCP/IP configuration (DHCP) and obtain all IP address infor-
mation automatically. See
Chapter 3.1.12.1, "Assigning an IP Address"
Remote command:
n/a
GPIB Address
Defines the analyzer's GPIB address. The address must be in the range between 0
and 30.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
SYSTem:COMMunicate:GPIB[:SELF]:DCLear:SUPPress
Remote Language
Selects the syntax of the R&S
ZNB/ZNBT's instrument control commands.
●
The DEFAULT language corresponds to the commands reported in this documen-
tation; see
Chapter 7.3, "SCPI Command Reference"
●
The ZVABT language ensures compatibility with network analyzers of the R&S
ZVA/B/T family. E.g. compared to the DEFAULT language, the command set does
not include
INITiate:CONTinuous:ALL
and
INITiate[:IMMediate]:ALL
.
The function of
INITiate:CONTinuous
and
INITiate[:IMMediate][:DUMMy]
is modified; refer to the remote control docu-
Chapter 7.3.9, "INITiate Commands"
●
The ZVR language ensures compatibility with network analyzers of the R&S ZVR
family. See also
Chapter 7.4, "R&S ZVR/ZVAB Compatible Commands"
●
PNA, ENA, HP8510, HP8720, HP8753 ... denote command sets for network ana-
lyzers from other instruments or manufacturers.
Note:
Remote Language settings other than DEFAULT are intended for remote control
of the analyzer. A mixed approach, with part of the instrument configuration defined via
the GUI, is possible but may cause unexpected results in some instances.