GUI Reference
R&S
®
ZNB/ZNBT
256
User Manual 1173.9163.02 ─ 38
The possible sweep stages and how they are displayed partly depend on the selected
sweep mode (see
Sweep Stage
Sweep Controller Display
in "Continuous" Sweep
Mode
in "Single" Sweep Mode
No ongoing sweep
Idle
Idle
Sweep is being prepared
Preparing
Preparing
Ongoing sweep
1)
Continuous
Sweeping
if #Sweeps = 1
Sweeping m/n
if n = #Sweeps > 1
Measurement results are being calculated
Continuous Calculation
Calculation
1)
This doesn't mean the VNA is actually sweeping; it may also wait for a trigger signal
5.3.3.2
Restart Manager Dialog
The "Restart Manager" dialog defines whether the active sweep mode ("Continuous" or
"Single") and the "Sweeps" value are valid for all channels in the active recall set or for
the active channel only.
Access:
CHANNEL – SWEEP > "Sweep Control" > "Restart Manager..."
This dialog is relevant / accessible in compatibility modes only (see
Related information
Refer to
Chapter 4.1.4, "Sweep Control"
Sweep Softtool