GUI Reference
R&S
®
ZNB/ZNBT
448
User Manual 1173.9163.02 ─ 38
External generators must be configured explicitly in the "External Generators" dialog
before they appear in the list. See also
Chapter 4.7.19, "External Generators"
Port
Selects the source for the stimulus signal. Depending on the state of the "Use Genera-
tor as Source" flag either analyzer ports ("Port 1" to "Port n") or external generators
("Gen 1", "Gen 2", ...) can be selected.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "<Ratio>"
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "<Ratio>"
(the parameter name <Ratio> can also contain the source port)
Detector
Selects the algorithm that is used to calculate the displayed measurement points from
the raw data.
Chapter 4.3.4.3, "Detector Settings"
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "<Ratio>"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "<Ratio>"
(the parameter name <Ratio> can also contain the detector)
[SENSe<Ch>:]SWEep:DETector:TIME
5.9.3 Wave Tab
Selects wave quantities as measured quantities.
Background information
Refer to
Chapter 4.3.4, "Wave Quantities and Ratios"
Meas Softtool