Command Reference
R&S
®
ZNB/ZNBT
1081
User Manual 1173.9163.02 ─ 38
[SENSe<Ch>:]SWEep:DETector:TIME
<DetectorTime>
Defines the detector observation time ratios and wave quantities if an "AVG Real Imag"
or an "AVG Mag Phase" detector is active. The detector is selected together with the
measured quantity (
CALCulate<Ch>:PARameter:MEASure
).
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time
Range:
0 s to 3456000 s
*RST:
0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1RMS'
Select the wave quantity a
1
for the default trace and activate the
RMS detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation:
See
[SENSe<Ch>:]SWEep:DWELl
<MeasDelay>
Defines the "Meas. Delay" before each partial measurement or the first partial mea-
surement (see
[SENSe<Ch>:]SWEep:DWELl:IPOint
). Setting a delay disables the
automatic calculation of the (minimum) sweep time (see
).
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasDelay>
Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep (
Range:
0 s to 13680 s
*RST:
0 s
Default unit: s
SCPI Command Reference