GUI Reference
R&S
®
ZNB/ZNBT
289
User Manual 1173.9163.02 ─ 38
During a measurement receiver calibration ("Meas. Receiver"), the following traces are
displayed:
●
The trace "a<m>(P<m>)" shows the (previously calibrated) power at the calibrated
reference plane (source port P<m>).
●
The trace "b<n>(P<m>)" shows the current power reading of the analyzer at the
calibrated receive port P<n> (source port P<m>).
After successful measurement receiver calibration, the "b<n>(P<m>)" trace should be
close to the "a<m>(P<m>)" trace. Due to the previous power calibration, both traces
should be close to the cal power.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:
RESult?
Port Overview
Shows all source ports together with the possible power calibrations. Either a source
power calibration (flatness calibration, "Power") or a measurement receiver calibration
("Meas. Receiver") can be performed at each analyzer port P1 ... PN.
●
If external generators are configured, they appear as additional source ports G1 ...
Gm in the port overview. A source power calibration is available at each generator
port.
●
If a "SMARTerCal" is active, (additional) receiver power calibrations are not
allowed and the R&S
ZNB/ZNBT only offers the possibility to perform "Source Flat-
ness" calibrations of the related test ports:
Chapter 4.5.7.3, "Combining SMARTerCal with Scalar Power Calibration"
on page 174 for details.
Select a "Power" or "Meas. Receiver" symbol to open the "Power Cal" dialog and to
perform a calibration sweep. A green checkmark indicates that the calibration data has
been acquired successfully.
See also
Remote command:
n/a, see
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled as soon as a new set of power calibration data has been acquired for either
a "Power" source or a "Meas. Receiver". The button applies all available source and/or
measurement receiver calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
Cal Softtool