Command Reference
R&S
®
ZNB/ZNBT
1110
User Manual 1173.9163.02 ─ 38
<CorrData>
Power correction values either in ASCII or block data format,
depending on the current
All numbers are interpreted as dB-values.
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
<PowerCalSettingType>[, <Index>]
Gets settings of the available power calibrations.
Suffix:
<Ch>
.
Number of the calibrated channel.
<Wv>
Number of the power calibration.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:
on page 1116 to query the number
of available power calibrations.
SCPI Command Reference