GUI Reference
R&S
®
ZNB/ZNBT
280
User Manual 1173.9163.02 ─ 38
Same Connector All Ports / Same Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some
multi-port calibration types, the port connector types must be equal, e.g. because they
require a Through standard with known characteristics.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Kit" dialog that allows to import a cal kit file. For back-
ground information see
Chapter 4.5.2.4, "Cal Kit Files"
Remote command:
Back
Go back to
.
Start
Start
.
Step 3: Calibration
Allows to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration type. On
"Apply" the R&S
ZNB/ZNBT calculates the system error correction (error terms) from
the measurement data of the standards and applies the result to the active channel.
The "Reduced Through" logic helps to keep the number of transmission measurements
as small as possible (see
Chapter 4.5.1.11, "Full n-Port Calibration with Reduced Num-
Cal Softtool