Command Reference
R&S
®
ZNB/ZNBT
996
User Manual 1173.9163.02 ─ 38
Example:
See
[SENSe<Ch>:]CORRection:COLLect:CONNection:
Manual operation:
See
"Same Connector All Ports / Same Gender All Ports"
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
<ConnectorType>
Qualifies whether the connector types at the test ports (but not their gender) are equal
or independent. Some calibration types require uniform port connector types.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<ConnectorType>
ALL | SINGle
ALL
–
equal (uniform) connector types. If the connector type at
one port is changed, the connector type at all other ports is
changed accordingly.
SINGle
–
independent (possibly non-uniform) connector types at
the ports.
*RST:
ALL
Example:
CORR:COLL:CONN:PORTS SING
Select independent connector types at the ports.
CORR:COLL:CONN1 N50MALE; CONN4 N75FEMALE;
CONN2?
Select independent connector types at ports 1 and 4. The con-
nector type at port 2 is not changed; the query returns
N50FEMALE
.
Manual operation:
See
"Same Connector All Ports / Same Gender All Ports"
[SENSe<Ch>:]CORRection:COLLect:DELete
[<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName>
Name of the calibration (string parameter) defined together with
the calibration type (
[SENSe<Ch>:]CORRection:COLLect:
ALL
- the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of
*RST:
-
SCPI Command Reference