Command Reference
R&S
®
ZNB/ZNBT
1054
User Manual 1173.9163.02 ─ 38
<ImaginaryPart>
Imaginary part of the port impedance. This part may be omitted
to define a real impedance.
Range:
- 10 MΩ to 10 MΩ
*RST:
Default reference impedance of the connector fam-
ily assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
Example:
PORT2:ZREF 52, 2
Specify a complex reference impedance of 52 Ω + j * 2 Ω for the
(physical) port no. 2.
CALC:TRAN:IMP:RNOR PWAV
Select renormalization of port impedances according to the
power waves theory.
Manual operation:
See
"Single Ended Mode / Common Mode / Differential Mode"
7.3.15.15
[SENSe:]POWer...
The
[SENSe:]POWer...
commands configure the optional receiver step attenuators
Chapter 4.7.15, "Receiver Step Attenuators"
on page 217) and the automatic/
adaptive gain control (AGC) of the R&S
ZNB/ZNBT.
See also
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
[SENSe<Ch>:]POWer:AGCMode:ACQuire
..................................................................... 1054
[SENSe<Ch>:]POWer:AGCMode:SAVE
.........................................................................1055
[SENSe<Ch>:]POWer:AGCMode<PhyPt>:MEASure
....................................................... 1055
[SENSe<Ch>:]POWer:IFGain<PhyPt>:MEASure
.............................................................1055
...............................................................................1055
................................................................................1056
[SENSe<Ch>:]POWer:GAINcontrol:ALL
......................................................................... 1057
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
................................................................... 1058
[SENSe<Ch>:]POWer:AGCMode:ACQuire
Starts an AGC learn sweep.
Suffix:
<Ch>
.
Channel number
Example:
Before you start, make sure channel 1 is fully configured (ports,
traces, ...).
*RST; :SENSe1:POWer:GAINcontrol:GLOBal MANual
Activate independent AGC settings for each drive port.
SENSe1:POWer:AGCMode:ACQuire;*WAI
Perform a Learn Sweep and wait until it has completed (asyn-
chronous command)
SENSe1:POWer:AGCMode:SAVE
Apply the acquired AGC settings to channel 1.
Usage:
Event
SCPI Command Reference