Command Reference
R&S
®
ZNB/ZNBT
1055
User Manual 1173.9163.02 ─ 38
Manual operation:
See
[SENSe<Ch>:]POWer:AGCMode:SAVE
Apply the static AGC (Automatic Gain Control) settings obtained during an AGC Learn-
ing Sweep.
Suffix:
<Ch>
.
Channel number
Example:
[SENSe<Ch>:]POWer:AGCMode:ACQuire
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]POWer:AGCMode<PhyPt>:MEASure
<Mode>
[SENSe<Ch>:]POWer:IFGain<PhyPt>:MEASure
<AGCMode>
These two equivalent commands activate manual gain control and select the IF gain in
the measurement channels (b-waves). The IF gain in the reference channel (a-waves)
is set automatically if not otherwise defined using
.
Note:
These commands are replaced by
and
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
. They are supported for compatibility
with FW versions < 1.5.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
Parameters:
<AGCMode>
LNOise | LDIStortion | AUTO
AUTO
–
automatic/adaptive gain control (AGC) according to the
RF input level.
LNOise
–
fixed, large IF gain, for low input levels.
LDIStortion
–
fixed, small IF gain, for high input levels.
*RST:
LDIS
Example:
*RST; :SENSe:POWer:IFG2:MEAS LNO; :SENSe:POWer:
IFG1:MEAS?
Optimize the received waves (measurement channel, b-waves)
at port 2 for small input levels and query the AGC settings for
the received waves at port 1. The response is
LDIS
.
[SENSe<Ch>:]POWer:ATTenuation
<arg0>[, <arg1>]
Sets an attenuation factor for the received waves. This command is available if at least
one of the
is installed.
SCPI Command Reference