![Rohde & Schwarz UPD Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/rohde-and-schwarz/upd/upd_application-note_1477885007.webp)
1GA16_1L.DOC
ROHDE & SCHWARZ
6
2.4.2 Preliminary Configuration of Setup File
The following modifications of the SPEAKER.SCO setup file are permissible:
•
selection of printer interface (SCREEN HARD COPY in OPTIONS panel),
•
defining a maximum generator voltage for protecting DUTs,
•
driving an external monitor (Extrn Disp in OPTIONS panel),
•
changing the X axis from log to linear (Spacing LIN POINTS in GENERATOR panel and X-AXIS-
>Spacing LIN in DISPLAY panel),
•
changing the Y1 axis from lin to log (TRACE A->Spacing LOG in DISPLAY panel), applies to
impedance measurements only,
•
selection of commands to be displayed in the status panel ("tick off"),
•
comments in plain text in the graphics display (DISPLAY panel) and info text (FILE panel)
•
comments in plain text for printout (dialog box of hardkey HCOPY),
•
defining tolerance curves/values and checks (LIMIT CHECK in DISPLAY panel).
The application software is tailored to the audio range, ie to the 25-kHz generator and a 22-kHz analyzer.
In the UPD instruments for higher frequencies may also be selected but in this case the following aspects
have to be considered:
•
the lower limit frequency of the analyzer is higher,
•
the resolution of the 8k FFT is only 41.67 Hz (100-kHz analyzer) or 125 Hz (300-kHz analyzer),
•
the measurement speed is lower.
With the preliminary configuration terminated, the setup has to be stored (as complete setup) under the
same name and may be assigned the attribute READ ONLY.
IMPORTANT
: Only the listed settings may be modified, as otherwise proper function of the software
cannot be guaranteed.
2
2.5
.5 Converting the Setup for Software Updates
Converting the Setup for Software Updates
If the device software is updated, the SPEAKER.SCO - and all other setups - may require to be
converted. This is done automatically when the setup is loaded but to avoid unnecessary delays during
loading, SPEAKER.SCO should be stored as converted setup. This can be done in two ways:
•
at the DOS level by calling up the batch converter:
⇒
UPD_CONV \SPEAKER\SPEAKER.SCO
or
⇒
UPL_CONV \SPEAKER\SPEAKER.SCO,
•
at the UPD/UPL level by loading and storing SPEAKER.SCO.
IMPORTANT: with setups set to READ ONLY, the "r" attribute has to be deleted at the DOS level:
⇒
ATTRIB -r \SPEAKER\SPEAKER.SCO