The
sinewave sweep
should be used whenever no significant echoes are expected, eg in the near field
or in anechoic chambers. Near-field measurements in the combined near/far-field measurement are
always performed with a sinewave sweep.
FIG 12: Measurement of transmission range using a sinewave sweep in an anechoic chamber
With the START softkey a level sweep or an FFT is started; the graphics display is in the unit selected
under 5.2.4, Selecting the Sound Pressure Unit.
SENSITIV has to be selected for an automatic determination of the transmission range after the
measurement. Full-screen display must be deactivated for the output to have enough space on the
screen. After the display of the frequency response characteristic the transmission range is calculated,
defined by a lower limit of -10 dB. When this limit is exceeded for more than 1/3 third (1/9 octave), the
limit of the transmission range is attained. Shorter drops are ignored. The transmission range determined
by the application program is indicated next to the graphics display.
Determination of the transmission range is started at the frequency specified under SNS FREQ. If the
level at this frequency is below -10 dBr, the calculation is stopped and a respective error message output.
5.2.7 Selecting the Combined Near/Far-Field Measurement
For the two sweep measurements, the far-field and the near-field can be configured separately,
determined with different test setups and frequency response curves can be combined.
The frequency range and the number of sweep points for the far field are determined under CFG FAR
(F7). If an additional near-field measurement is not performed, the complete frequency response
measurement is carried out with these settings.