![Rohde & Schwarz UPD Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/rohde-and-schwarz/upd/upd_application-note_1477885031.webp)
1GA16_1L.DOC
ROHDE & SCHWARZ
30
5.3.3 Linearity Departure and Group Delay
Assessing the phase linearity from the sweep diagram is problematic because a range of up to 180000°
(1024
*
180°) has to be displayed in the Y direction with linear scaling. A better overview is provided by
two alternative display modes, linearity departure and group delay. They are selected with softkeys F9 to
F11 at softkey level 3 or 4 before or after the sweep.
•
PHASE (F9) displays the sweep results unchanged with absolute phase values,
•
LINEAR (F10) displays the deviations from an ideal straight line. The slope of this line is determined
between 2 frequencies to be entered in REF AREA.
•
GRP DEL (F11) displays the group delay versus frequency.
FIG 14 Group delay derived from the phase characteristic
5
5.4
.4 Polarity
Polarity
Correct polarity of crossover units, loudspeaker arrays and systems becomes important if partly or
completely identical signals are applied to several loudspeakers. While an incorrectly poled mono
broadband loudspeaker would hardly be noticed, superposition of the 180°-phase-shifted sound waves of
two stereo loudspeakers produces audible cancellations. This also applies to incorrectly poled single
systems in a multiple loudspeaker, where cancellations occur also in the transition range of the crossover
unit.
5.4.1 Measurement Function POLARITY for Electrical Systems
UPD and UPL provide a measurement function and a corresponding generator function for measuring the
polarity of electrical systems (cables, crossover units). In this case a (DC-loaded) sine2 pulse is generated
and the measured polarity is displayed as "+ 1 Pol" (correct polarization) or "- 1 Pol" (wrong polarization).
However, this method cannot be used for loudspeakers.
Содержание UPD
Страница 38: ...1GA16_1L DOC ROHDE SCHWARZ 37 ...