5.3.1 Test Setup
The test setup is the same as for sound pressure measurements (see section 5.2.1), but the use of a
power amplifier is not obligatory. In a setup without power amplifier, the generator outputs are connected
in parallel to the loudspeaker and to the XLR input of the UPD/UPL analyzer (channel 1).
The standard microphone is connected to the XLR input (channel 2) of the UPD analyzer (if required via a
microphone amplifier).
FIG 13: Test setup for phase measurement
5.3.2 Frequency Response of Phase
The phase linearity should be checked in a linear frequency sweep where the phase is measured between
the loudspeaker input and the generated sound wave. In the case of a linear phase response, a line with a
negative slope would ideally be obtained. In order to obtain a continuous phase response without 360°
shifts, the sweep points displayed in the range from 0 to 360° have to be further processed. The
algorithm of the application software requires in addition that the phase values measured at two adjacent
sweep points differ by less 180°. At a defined microphone distance the signal travel time is
t s
d m
/
/
=
330
The requirement
ϕ
ϕ
2
1
2
1
360
180
−
=
°⋅
−
⋅ <
°
(
)
f
f
t
yields for the frequency increment
(
) /
/
f
f
Hz
t s
2
1
1
2
−
<
⋅