Installation and Commissioning
R&S
®
TS7124AS
37
Instructions Handbook 1179.2948.02 ─ 01
To activate the chamber
Connecting the chamber to power activates it. It does not have a separate [ON
/
OFF]
switch.
1.
CAUTION!
Initial door movement can cause injury. When you connect the chamber
to power, the door can move abruptly.
To prevent unexpected movement of the door upon first activation, perform the fol-
lowing safety measure.
Pull the door fully open.
2. Connect the DC plug of the power supply unit (included in the delivery) to the
24
V
DC power supply connector, see
"To prepare the power connection"
shows the resulting connections.
The chamber is activated.
The pneumatic system moves the door to its fully opened position, while the status
LED in the push-button switch unit remains turned off.
When you push the button again or send the
command, the door closes
and the status LED turns green.
To configure the LAN network
Only a
► Configure the LAN network by using the commands described in
"Remote Configuration Commands"
The default IP address is 192.168.178.41, port 5000.
6.6
Testing Safety Systems
automatic emergency deactivation
mechanism for proper functioning
as follows:
1. Activate the chamber according to
2. Open the door according to
Chapter 7.3.2, "Push-Button Door Operation"
3. Hold a flat obstacle across the edge of the door opening.
We recommend using a flat piece of wood or similar material, 1
cm to 2
cm thick.
Alternatively, use the handle of a screwdriver.
4. Close the door.
5. Verify that the door's automatic emergency deactivation mechanism stops the door
when it hits the obstacle, as described in
If the test fails, see
on page 38, and proceed as described in
.
6.
WARNING!
Risk of personal injuries. See
Testing Safety Systems