Function Description
R&S
®
TS-PIO2
23
User Manual 1506.7208.12 ─ 06
Figure 5-10: Ground relay
The module is operated ground-free in its basic state. This state can be changed using
the function
rspio2_ConfigureGround
.
Please note that function
rspio2_DisconnectAll
does not open the ground relays!
For technical reasons, a non-switched R&S TS-PIO2 module (no connection of signals
to the front side connector or analog bus) is automatically grounded through the
ground relay. This relay is automatically opened again before new switching is per-
formed. This applies if the R&S TS-PIO2 module is configured groundfree.
5.1.4 Using sense lines
To compensate for voltage drops in the power supply to the external load, the exten-
ded channels (CH4, CH8, CH12 and CH16) of R&S TS-PIO2 can be set to external
sensing. Two additional lines directly to the test object are required for this purpose.
The measured difference in voltage on these lines is automatically controlled to the tar-
get voltage by R&S TS-PIO2.
The sense lines on the front side connector (CHx_SHI und CHx_SLO) are switched
using function
rspio2_ConfigureRemoteSensing
.
5.1.5 Adjusting current limiting
The extended channels (CH4, CH8, CH12 and CH16) of the R&S TS-PIO2 make it
possible to adjust current limiting. The set value is independent of the mode of a chan-
Function description of the R&S TS-PIO2 module