Putting into Operation
R&S
®
RT-ZS60
20
User Manual 1418.7342.02 ─ 01
Fig. 2-2: Offset compensation voltage and dynamic range
To set offset compensation on the front panel:
1. Press the Vertical Position knob until the "Offset" setup is shown on the display.
2. Turn the Vertical Position knob.
To automatically set the offset compensation voltage using the menu:
1. On the "Vertical" menu, select "Probe Setup".
2. Select the channel and tap "Set offset to mean".
For more details on setting the offset compensation voltage, refer to the "User
Manual" of the R&S oscilloscope.
Additionally, the probe offers automatic offset compensation by means of the micro
button. A single push on the button compensates the DC component of the mea-
surement signal which is particularly helpful during measurements of input signals
with different DC offsets (see also
Offset Compensation