Instrument functions
R&S
®
NGP800
65
User Manual 5601.5610.02 ─ 09
–
"Off": Output of the selected channel is turned off after performing the
QuickArb function.
–
"Hold": Last voltage and current values output at the instrument.
10. Select [Back] key to return to channel menu dialog.
11. Activate the "Arbitrary" menu item to enable the QuickArb function.
The R&S
NGP800 enables the QuickArb function and displays the "Arbitrary" icon
on the selected channel status bar information.
Arbitrary editor
The "Arb Editor" dialog allows you to edit the arbitrary profile ("Voltage", "Current",
"Time", "Interpolate" status, "Rep." and "End Behavior"). To view or open the list of
available arbitrary files, select to open the arbitrary file.
Figure 6-21: Arbitrary editor dialog
1. Configure the "Arb Editor" with the required voltage, current and duration.
The R&S
NGP800 displays the on-screen keypad for data entry.
2. Confirm values with the unit keys.
3. Select the interpolation checkbox to toggle on/off the interpolation function on the
arbitrary data.
4. Select the or icon to add or delete the arbitrary data from the dialog.
●
: A new row of data is added at the end of the table. It is initially filled with the
data of the currently selected row.
●
: Removes the currently selected row from the table.
5. Set the "Rep" to configure repetition cycle for the arbitrary data.
By default, the repetition cycle is set to infinity.
6. Set the "End Behavior" to handle the way to end the automation of the QuickArb
function.
Advanced features