Remote control commands
R&S
®
NGP800
111
User Manual 5601.5610.02 ─ 09
Example:
SYSTem:DATE 2018, 10, 15
SYSTem:DATE?
-> 2018, 10, 15
Returns the system date.
SYSTem:KEY:BRIGhtness
<brightness>
Sets or queries the front panel key brightness.
Parameters:
<brightness>
Sets the key brightness.
Range:
0.0 to 1.0
Increment: 0.1
*RST:
1.0
Example:
SYSTem:KEY:BRIGhtness 1.0
SYSTem:KEY:BRIGhtness?
-> 1.0
Returns key brightness value: 1.0.
SYSTem:INTerface?
Queries the available system interface.
Usage:
Query only
SYSTem:INTerface:GPIB?
Queries the GPIB interface information.
Usage:
Query only
SYSTem:LOCal
Sets the system to front panel control. The front panel control is unlocked. If the front
panel control was locked with the SCPI command
SYSTem:RWLock
, the message box
of the locked front panel on the instrument display will be disappeared.
Usage:
Event
SYSTem:REMote
Sets the system to remote state. The front panel control is locked. By pushing the soft-
key button [*] key, the front panel control will be activated.
Usage:
Event
SYSTem:RWLock
Sets the system to remote state. The front panel control is locked and a message box
is shown on the instrument display. You are only able to unlock the front panel control
via SCPI command
SYSTem:LOCal
.
System settings commands