Instrument functions
R&S
®
NGP800
61
User Manual 5601.5610.02 ─ 09
Table 6-2: Trigger out parameters and conditions
Trigger out
parameters
Trigger conditions
Description
Channel
"--", "Ch 1", "Ch 2", "Ch 3", "Ch 4"
Output channel selected to monitor for trig-
ger conditions.
If "--" is selected, no channel is selected for
trigger response.
Mode
Output
Output the selected logic level when the out-
put is turned on at the selected channel.
Fuse
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
Chapter 6.4.1, "Overcurrent protection
Operation mode
●
"CC": Output the selected logic level
when the selected channel operates in
the CC mode. See
●
"CV": Output the selected logic level
when the selected channel operates in
the CV mode. See
Voltage Level, "Vset"
>= "set
value"
Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set voltage level.
Current Level, "Iset"
>= "set
value"
Output the selected logic level when the cur-
rent level of the selected channel is greater
or equal to the set current level.
Power Level, "Plevel"
>= "set
value"
Output the selected logic level when the
power level of the selected channel is
greater or equal to the set power level.
Critical event
"OVP"
Output the selected logic level when the
selected critical event ("OVP", "OPP",
"OTP") occurs on the selected channel. See
"OPP"
"OTP"
"Logging"
Output the selected logic level when the log-
ging is enabled.
Active Level
High
Set the logic level of the trigger out signal.
Low
Digital trigger I/O