Remote control commands
R&S
®
NGP800
167
User Manual 5601.5610.02 ─ 09
State
Descriptions
0
Idle
1
Busy
2
Waiting
12
Voltage adjustment completed
13
Current adjustment completed
16
Successful channel adjustment
17
Failed channel adjustment
Example:
CAL:STAT?
->
12
Voltage adjustment is successful.
Usage:
Query only
CALibration:TEMPerature?
Returns the temperature of selected channel.
Usage:
Query only
CALibration:USER
Starts the channel adjustment process.
Usage:
Event
CALibration:VOLTage:DATA
<arg0>
Sets the DMM reading after setting the output voltage level in channel adjustment
process.
Parameters:
<voltage>
Measured value from DMM.
CALibration:VOLTage:UMAX
Sets the output voltage to high value 100 % of Vmax during voltage adjustment.
Usage:
Event
CALibration:VOLTage:UMIN
Sets the output voltage to low value 1 % of Vmax during voltage adjustment.
Usage:
Event