SCPI Commands HMF series
Remote Control
23
Command Reference
SCPI Programmers Manual
OUTPut:LOAD {<Impedance in Ohm> | TERMinated | INFinity}
Defines the instrument output load.
Parameters:
<Load>
Numeric value in Ohm.
TERMinated
: 50
Ω
termination.
INFinity
:
Open circuit (without any load).
*RST:
INF
OUTPut:LOAD?
Queries the type of the instrument output load.
Return values
:
Numeric value in
Ω
| INF
OUTPut:POLarity {NORMal | INVerted}
Defines the instrument output polarity.
Parameter:
<Type>
NORMal | INVerted
NORMal
:
Default instrument output polarity.
INVerted
:
The instrument output polarity will be inverted.
NOTICE
The OUTPut:POLarity command is only valid for the pulse and arbitrary function.
OUTPut:POLarity?
Queries the type of the instrument output polarity.
Return values
:
NORM | INV
OUTPut:OFFSet {OFF | ON | 0 | 1}
Activates or deactivates the output offset voltage.
Parameters:
<State>
ON | 1
:
The offset voltage will be activated.
OFF | 0
:
The offset voltage will be disabled.
*RST: OFF | 0