56
Instrument Functions
In the SETUP menu, use the arrow keys
7
to se-
lect the GUARD function, press the knob
6
(editing mode)
and select the desired setting via knob. Press the knob
again to confirm the selection.
The HM8118 GUARD function is not comparable to the
4TP function (= Four Terminal Pair) of other measuring in-
strument manufacturers. For the 4TP function, the mea-
surement current is returned through the test lead shield.
The electromagnetic radiation of the supply and return
conductor nearly override each other which for the most
part resolves the issue of electromagnetic coupling. This
does not work for the Kelvin test lead provided with the
R&S®HM8118, as this is not properly converted (the shields
would have to be short-circuited preferably close to the
test point). The HM8118 uses a 5 terminal configuration /
5T and does not support the 4TP function.
7.1.11 Deviation DEV_M
You can use the DEV_M function to turn on or off (OFF)
the display of the measurement deviation of the main dis-
play (Main) in Δ % (percent) or Δ ABS (absolute) as applied
to the reference value REF_M. In the SETUP menu, use the
arrow keys
7
to select the DEV_M function to
set the display for the measurement deviation, press the
knob
6
(editing mode) and select the desired setting via
knob. Press the knob again to confirm the selection. For
more information about the measurement value deviation,
see chapter 6.
7.1.12 Reference REF_M
You can use the REF_M function to save the measure-
ment value as a reference value in the reference memory
M (Main). You can choose one of the following as unit
for the measurement value: H, mH, µH, nH, F, mF, µF, nF,
pF, Ω, mΩ, kΩ, MΩ, or S, kS, mS, µS, nS, pS. In the SE
-
TUP menu, use the arrow keys
7
to select the
REF_M function to set the reference value, press the knob
6
(editing mode) and select the desired reference value
via knob. By pressing the knob again, you can activate the
manual value input via numeric keypad. A value input win-
dow will be opened. You can use the numeric keys to en-
ter a value. After entering the value via keypad, confirm the
entry by pressing the ENTER key or by pressing the knob
again. As long as this field is activated, you can also use
the TRIG key
27
to accept the value of the DUT (= Device
Under Test). For more information about the reference va-
lue, see chapter 6.
7.1.13 Deviation DEV_S
You can use the DEV_S function to turn on or off (OFF) the
display of the secondary value display (Sub) in Δ % (per
-
cent) or Δ ABS (absolute) as applied to the reference value
It is recommended to use the GUARD function if measurement
adapters with high capacity (e.g. R&S®HZ184) are used. If the
DUT exhibits impedances of more than 25kΩ at frequencies of
more than 100kHz, is is also recommended to use the GUARD
function.
REF_S. In the SETUP menu, use the arrow keys
7
to select the DEV_S function to set the display for the
measurement deviation, press the knob
6
(editing mode)
and select the desired setting via knob. Press the knob
again to confirm the selection. For more information about
the measurement value deviation, see chapter 6.
7.1.14 Reference REF_S
You can save a measurement value of the dissipation fac-
tor or the quality factor (quality) as reference value in the
reference memory S. You can choose one of the following
as unit for the measurement value: Ω, mΩ, kΩ, MΩ, S,
kS, mS, µS, nS, pS or °. In the SETUP menu, use the ar-
row keys
7
to select the REF_M function to
set the reference value, press the knob
6
(editing mode)
and select the desired reference value via knob. By pres-
sing the knob again, you can activate the manual value in-
put via numeric keypad. A value input window will be ope-
ned. You can use the numeric keys to enter a value. After
entering the value via keypad, confirm the entry by pres
-
sing the ENTER key or by pressing the knob again. As long
as this field is activated, you can also use the TRIG key
27
to accept the value of the DUT (= Device Under Test). For
more information about the reference value, see chapter 6.
7.1.15 CONSTANT VOLTAGE CST V
The CST V function allows you to turn the constant voltage
(AC) on (ON) or off (OFF). Due to the source resistance,
some test require the use of a specific measurement vol
-
tage which is not possible with the regular source resis-
tance of the respective measurement range. In the SETUP
menu, use the arrow keys
7
to select the CST V
function to activate the constant voltage, press the knob
6
(editing mode) and select the desired setting via knob.
Press the knob again to confirm the selection.
If the constant voltage is activated (ON), the source
resistance is preset to 25 Ω. The voltage applied to the
component will be nearly constant for all components
whose impedance is substantially greater than 25 Ω. If the
constant voltage mode is activated for the bridge, the
measurement range changes (depending on the
impedance of the connected component) to prevent
overloading the bridge. However, the accuracy is reduced
by the factor of 2 in the constant voltage mode. The
following table shows the impedance measurement
ranges when the constant voltage mode is activated (CST
V ON):
The constant voltage (CST V function) must be switched on for
measurements with BIAS bias current or external BIAS preload.
Measure-
ment Range
Source
Resistance
Component
Impedance
1
25 Ω
10.0 µΩ to 3.0 Ω
2
25 Ω
3.0 Ω to 100.0 Ω
3
25 Ω
100.0 Ω to 1.6 kΩ
4
25 Ω
1.6 kΩ to 25.0 kΩ
5
25 Ω
25.0 kΩ to 2.0 MΩ
6
25 Ω
2.0 MΩ to 100.0 MΩ